PSW Series Programming Manual
70
[SOURce:]CURRent:PROTection:STATe
Set
Query
Description
Turns OCP (over-current protection) on or off.
Syntax
[SOURce:]CURRent:PROTection:STATe {0|1|OFF|ON}
Query Syntax
[SOURce:]CURRent:PROTection:STATe?
Parameter/Return
0
<NR1> Turns the buzzer off.
OFF
Turns the buzzer off.
1
<NR1> Turns the buzzer on.
ON
Turns the buzzer on.
Return parameter
<Bool>
Returns bleeder resistor status (0 or 1).
Example
SOUR:CURR:PROT:STAT OFF
Turns OCP off.
[SOURce:]CURRent:SLEW:RISing
Set
Query
Description
Sets or queries the rising current slew rate. This is
only applicable for CC slew rate priority mode.
Syntax
Query Syntax
[SOURce:]CURRent:SLEW:RISing {<NRf>|MIN|MAX}
[SOURce:]CURRent:SLEW:RISing? [MIN|MAX]
Parameter/Return
<NRf>
0.01A/s~72.00A/s (PSW 30-36)
0.1A/s~144.0A/s (PSW 30-72)
0.1A/s~216.0A/s (PSW 30-108)
0.01A/s~27.00A/s (PSW 80-13.5)
0.01A/s~54.00A/s (PSW 80-27)
0.01A/s~81.00A/s (PSW 80-40.5)
0.01A/s~14.40A/s (PSW 160-7.2)
0.01A/s~28.80A/s (PSW 160-14.4)
0.01A/s~43.20A/s (PSW 160-21.6)
0.001A/s ~ 9.000A/s (PSW 250-4.5)
0.01A/s ~ 18.00A/s (PSW 250-9)
0.01A/s ~ 27.00A/s (PSW 250-13.5)
0.001A/s ~ 2.880A/s (PSW 800-1.44)
0.001A/s ~ 5.760A/s (PSW 800-2.88)
0.001A/s ~ 8.640A/s (PSW 800-4.32)
MIN
Minimum rising current slew rate.
MAX
Maximum rising current slew rate.