REMOTE CONTROL 
33 
  MAX 
Maxium value for the present range. 
Return parameter  <NRf>  Returns the voltage and  current. 
Example APPL 5.05,1.1 
Sets the voltage and current to 5.05V and 1.1A. 
Query Example  APPL? 
+5.050, +1.100 
Returns voltage (5.05V) and current (1.1A) setting. 
Display Commands 
 
 
DISPlay:MENU[:NAME] ....................................................33
DISPlay[:WINDow]:TEXT:CLEar ..................................... 34
DISPlay[:WINDow]:TEXT[:DATA] ..................................  34
DISPlay:BLINk ...................................................................... 34
 
DISPlay:MENU[:NAME] 
Set
Query
 
Description 
The DISPlay MENU command selects a screen 
menu or queries the current screen menu.   
Syntax 
Query Sytax 
DISPlay:MENU[:NAME] <NR1> 
DISPlay:MENU[:NAME]? 
Parameter/ 
Return parameter 
<NR1> 
Description 
0 
Measurement-Voltage / Measurement-
Current  
  1 
Measurement-Voltage / Measurement-Power  
  2 
Measurement-Power / Measurement-Current  
  3 
Set Menu  
  4 
OVP / OCP Menu  
  5~99 
Not Used.  
  100~199 
F-00~99 Menu.  
Example DISP:MENU:NAME 0 
Sets the display to the Voltage/Current display screen.