REMOTE CONTROL
41
STATus:QUEStionable:CONDition
Query
Description
Queries the status (bit sum) of the Questionable
Status register. This query will not clear the
register.
Query Syntax STATus:QUEStionable:CONDition?
Parameter <NRf>
0~32767
Return parameter <NR1>
0~32767
STATus:QUEStionable:ENABle
Set
Query
Description
Sets or queries the bit sum of the Questionable
Status Enable register.
Syntax
Query Syntax
STATus:QUEStionable:ENABle <NRf>
STATus:QUEStionable:ENABle?
Parameter <NRf>
0~32767
Return parameter <NR1>
0~32767
STATus:QUEStionable:PTRansition
Set
Query
Description
Sets or queries the bit sum of the positive
transition filter of the Questionable Status register.
Syntax
Return Syntax
STATus:QUEStionable:PTRansition <NRf>
STATus:QUEStionable:PTRansition?
Parameter <NRf>
0~32767
Return parameter <NR1>
0~32767
STATus:QUEStionable:NTRansition
Set
Query
Description
Sets or queries the negative transition filter of the
Questionable Status register.
Syntax
Query Syntax
STATus:QUEStionable:NTRansition <NRf>
STATus:QUEStionable:NTRansition?