REMOTE CONTROL
55
Parameter 0
Output off at power up
OFF
Output off at power up
1
Output on at power up
ON
Output on at power up
Return Parameter 0
Output off at power up
1
Output on at power up
SYSTem:COMMunicate:ENABle
Set
Query
Description
Enables/Disables LAN, GPIB or USB remote
interfaces as well as remote services (Sockets, Web
Server).
Syntax
Query Syntax
SYSTem:COMMunicate:ENABle <mode>,<interface>
SYSTem:COMMunicate:ENABle? <interface>
Parameter <mode>
OFF
Turns the selected mode off.
0
Turns the selected mode off.
ON
Turns the selected mode on.
1
Turns the selected mode on.
<interface>
GPIB
Select GPIB
USB
Select USB
LAN
Select LAN
SOCKets
Select Sockets
WEB
Select the web server
Return Parameter 0
The selected mode is off.
1
The selected mode is on.
Example SYST:COMM:ENAB 1,USB
Turns the USB interface on.
Query Example SYST:COMM:ENAB? USB
1
Queries the USB state, returns 1 (USB is on).