REMOTE CONTROL
49
Syntax TRIGger:TRANsient[:IMMediate]
TRIGger:TRANsient:SOURce
Set
Query
Description
Sets or queries the trigger source for the transient
system.
Syntax
Query Syntax
TRIGger:TRANsient:SOURce {BUS | IMMediate}
TRIGger:TRANsient:SOURce?
Parameter/Return
BUS
Internal software trigger. Waits for the
*TRG (or IEEE 488.1 “get” group
execute trigger) command to start the
trigger.
IMMediate
Starts the trigger immediately.
(default)
Example TRIG:TRAN:SOUR BUS
Sets the trigger source as BUS.
TRIGger:OUTPut[:IMMediate]
Set
Description
Generates a software trigger for the output trigger
system.
Syntax TRIGger:OUTPut[:IMMediate]
TRIGger:OUTPut:SOURce
Set
Query
Description
Sets or queries the trigger source for the output
system.
Syntax TRIGger:OUTPut:SOURce {BUS | IMMediate |
EXTernal}
Query Syntax TRIGger:OUTPut:SOURce?
Parameter/Return
BUS
Internal software trigger. Waits for the
*TRG (or IEEE 488.1 “get” group
execute trigger) command to start the
trigger.
IMMediate
Starts the trigger immediately.
(default)