EasyManua.ls Logo

GW Instek PSW 80-27 - Page 43

GW Instek PSW 80-27
87 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
43
Source Commands
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] 43
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] .. 44
[SOURce:]CURRent:PROTection[:LEVel] ....................... 44
[SOURce:]CURRent:SLEW:RISing .................................... 45
[SOURce:]CURRent:SLEW:FALLing ................................ 45
[SOURce:]RESistance[:LEVel][:IMMediate]
[:AMPLitude] .......................................................................... 46
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] 46
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] . 47
[SOURce:]VOLTage:PROTection[:LEVel] ....................... 47
[SOURce:]VOLTage:SLEW:RISing .................................... 47
[SOURce:]VOLTage:SLEW:FALLing ............................... 48
[SOURce:]CURRent[:LEVel][:IMMediate]
[:AMPLitude]
Set
Query
Description
Sets or queries the current level in amps. For
externally set current levels (from the analog
control connector) the set current level is returned.
Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
{<NRf>|MIN|MAX}
Query Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
{MIN|MAX}
Parameter/Return
<NRf>
0~105% of the rated current output level.
MIN
Minimum current level.
MAX
Maximum current level.
Example SOUR:CURR:LEV:IMM:AMPL? MAX
37.800
Returns the maximum possible current level in amps.

Table of Contents

Other manuals for GW Instek PSW 80-27

Related product manuals