Figure 3 Product components
1 FL1500 flow logger 3 AC power cable with country-specific connector
2 Communications cable, USB A to B
Installation
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Installation guidelines
Do not install the instrument in a location that receives direct exposure to sunlight, ultraviolet
radiation (UV), severe weather or next to a heat source.
Make sure that there is sufficient clearance around the instrument to make connections.
Install the instrument in an environmental enclosure or protective cover when installed outdoors.
Mechanical installation
Mounting
This instrument is rated for an altitude of 2000 m (6562 ft) maximum. Although the use of this
equipment above the 2000 m altitude does not show any substantial safety concern, the
manufacturer recommends that users with concerns contact technical support.
Attach the instrument to a wall, a rail or a pipe
1
with a diameter of 20 to 50 mm (0.75 to 2.0 inch).
Refer to Figure 4 and Figure 5. Make sure that the wall mounting is able to hold 4 times the weight of
the equipment.
8
English