Table 15 Main menu options (continued)
Option Description Option Description
Calibration
Calibrates the installed sensors.
Review Data
Shows the measurement data (the data
log).
Diagnostics
Shows the status screen, event
log, alarm log, sensor data and
internal I/O data. Has a test for
the keypad and display.
General settings
Shows the instrument details (e.g.,
serial number). Configures the general
instrument and display settings. Erases
data, sets security, exports/imports from
a USB memory stick, restores defaults.
Status indicators
The indicator light shows the status of the instrument. Refer to Table 16.
Table 16 LED status indicators
LED color Description
Green Flashes during normal operation.
Red Flashes when one or more alarms are active.
Blue Flashes when there is a communication problem with the IO9000 module or with a sensor with
logged channels.
Orange Flashes when the internal memory battery is low. Contact factory service immediately.
Operation
Configuration options—Instrument or PC
N O T I C E
Damage to the logger and PC can occur if the PC is connected to the incorrect port on the logger. Connect the
PC only to the USB B port on the logger.
Use the keypad on the instrument or a PC with the applicable software to configure the instrument.
This document contains instructions to configure the instrument from the keypad.
To use a PC for configuration, refer to the documentation for the applicable software to configure the
instrument. When the configuration is complete, import the configuration file directly from the PC or
from a USB memory stick. Refer to Figure 1 on page 6 to connect a PC or USB memory stick to the
instrument. Refer to Import or export the instrument settings on page 38 to import the configuration
file.
Configure the general settings
Use the general settings menu to get instrument information, change the display settings, erase data,
set security, export/import from a USB port, restore defaults.
1. Push MENU.
2. Select General Settings.
3. Select an option.
English
27