3. Select an option:
Option Description
Internal I/O Configures the inputs, outputs and relays in the instrument.
External I/O Configures the inputs, outputs and relays in a connected IO9004 module. Select External
I/O>IO9004>Enable.
4. Select the input or output option:
Option Description
Analog inputs Enables or disables the analog inputs. Selects the type of measurement, the 0–20 mA or
4–20 mA scale and the values for the minimum and maximum signal. Calibrates the
minimum and maximum signal (optional).
Analog outputs Enables or disables the analog outputs. Selects the measurement channel, the 0–20 mA or
4–20 mA scale and the values for the minimum and maximum signal. Sets the power mode
to internal or external. Sets a transfer value for use during maintenance tasks. Calibrates
the minimum and maximum signal (optional).
AC relays Enables or disables the relays.
Digital outputs Enables or disables the digital outputs.
Digital inputs Enables or disables the digital inputs. Selects the trigger value options: low to high or high
to low. Sets the voltage threshold to internal or external.
Configure flow pacing for connected samplers
Use flow pacing to set up an attached automatic sampler to collect samples at specified flow
intervals. Make sure that the wires from the sampler cable connect to the auxiliary port in the logger.
Note: The sampler can also collect a sample during alarm conditions if the Trigger Sampler action is set for a
channel alarm.
1. Push MENU.
2. Select Hardware Setup>Sampler.
3. Select the options.
Option Description
Source Selects the flow channel to measure the flow for sampler pacing.
Interval Sets the volume interval that sends a flow pulse to the sampler. For example, a sampler can collect
a sample for each 100 gallons of flow.
Unit Selects the flow units. Options: gal (gallons, default), ltr (liters), af (acre feet), ft
3
(cubic feet), m
3
(cubic meters).
Configure network communications
Configure the instrument for network communications when an RS485 cable is installed in the
instrument.
1. Push MENU.
2. Select Hardware Setup>Communication>RS485.
3. Select an option to change the address, baud rate or parity. The default baud rate is 115200 and
the default parity is none.
Data management
View data
View the measurement data on the display screen or on a PC with the applicable software. This
document contains instructions to view the measurement data from the instrument.