Connect to optional devices
Refer to the steps that follow to connect the optional devices from the manufacturer.
1. Remove the power to the instrument.
2. Open the access door. Refer to Open the access door on page 12.
3. Put the cable through a strain relief fitting.
4. Install each wire in the terminal block. Refer to Figure 13 and the wiring information for the
applicable device:
Option Description
AS950 sampler Connect the sampler to the AUXILIARY terminal block with the sampler half cable. Refer to
Table 5.
Rain gauge Connect the rain gauge to the RAIN terminal block. Refer to Table 6.
IO9004 module Connect the module to the IO9004 terminal block. Refer to Table 7.
Totalizer Connect the mechanical totalizer to the TOTALIZER terminal block. Refer to Table 8.
5. Tighten the cable strain relief.
6. Install the access door.