2—19
User-Entered Calibration, continued
Note:
Press the READ/ENTER key to move from digit to digit. Use the ZERO/SCROLL
key to change the number.
8. After completing entry of the blank value, press the READ/ENTER key. The
display will show “A0”.
9. Insert the reagent blank or deionized water into the cell holder. Cover the
blank with the instrument cap.
10. Press the
READ/ENTER key. The meter will measure and display the
absorbance value for “S0”.
11. Remove the sample blank. Press the
ZERO/SCROLL key. “S1” will appear.
Press the
READ/ENTER key, then enter the first standard value.
Note: Press the READ/ENTER key to move from digit to digit. Use the ZERO/SCROLL
key to change the number.
12. After completing entry of the first standard value, press the READ/ENTER
key. The display will show “A1”.
13. Insert the first reacted standard solution into the cell holder. Cover the
prepared standard with the instrument cap.
5957088.book Page 19 Friday, March 22, 2013 3:54 PM

Other manuals for Hach POCKET COLORIMETER II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach POCKET COLORIMETER II and is the answer not in the manual?

Hach POCKET COLORIMETER II Specifications

General IconGeneral
Measurement MethodColorimetric
DisplayLCD
Operating Temperature0 to 50 °C
Storage Temperature-40 to 60 °C
Waterproof RatingIP67
Weight0.23 kg
Light SourceLED
RangeVaries by parameter
WavelengthVaries by parameter
Battery LifeApproximately 2000 tests

Related product manuals