1—45
Chlorine, Free and Total, HR, continued
Accuracy Check
Standard Additions Method
a. Use the ampule breaker to snap the neck off a high range Chlorine
Standard Solution Ampule, 50–70 mg/L Cl
2
.
b. Use a TenSette
®
pipet to add 0.1, 0.2, and 0.3 mL of standard to three
5-mL samples. Swirl gently to mix.
c. Analyze each sample as described in the procedure. Each 0.1 mL of
standard will cause an incremental increase in chlorine. The exact
value depends on the concentration of the ampule standard. Check the
certificate enclosed with the ampules for calculation of the expected
increase in the chlorine concentration.
Standard Solution Method
Standard solutions for chlorine are difficult and time-consuming to prepare.
Errors can occur if attention to detail is not addressed during preparation of the
standards. The calibration curve is prepared under rigorous analytical
laboratory conditions. Use the factory calibration for most normal testing.
5957088.book Page 45 Friday, March 22, 2013 3:54 PM

Other manuals for Hach POCKET COLORIMETER II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach POCKET COLORIMETER II and is the answer not in the manual?

Hach POCKET COLORIMETER II Specifications

General IconGeneral
Measurement MethodColorimetric
DisplayLCD
Operating Temperature0 to 50 °C
Storage Temperature-40 to 60 °C
Waterproof RatingIP67
Weight0.23 kg
Light SourceLED
RangeVaries by parameter
WavelengthVaries by parameter
Battery LifeApproximately 2000 tests

Related product manuals