User-Entered Calibration, continued
• To select Edit from the User menu, press ZERO/SCROLL and READ/ENTER.
• Once in the CAL or Edit option, press the READ/ENTER key to navigate
through each option.
Note: Press ZERO/SCROLL to quickly scroll through each option.
Calibration Procedure Using Prepared Standards
Note: Deionized water or a reagent blank can be used to zero during the calibration
procedure. Calibrations generated with deionized water as the zero will give
less accurate results if the reagent blank is significantly more turbid or colored
than deionized water. Use the deionized water or the reagent blank as the zero
concentration point (S0) in the following calibration procedure.
1. Turn on the instrument and select the range to be calibrated. An arrow at
the top of the display will point to the selected range. To change ranges,
press the
MENU key, then use the READ/ENTER key to toggle between
ranges 1 and 2. Press
MENU again to return to measurement mode.