2—17
User-Entered Calibration, continued
To select Edit from the User menu, press ZERO/SCROLL and READ/ENTER.
Once in the CAL or Edit option, press the READ/ENTER key to navigate
through each option.
Note: Press ZERO/SCROLL to quickly scroll through each option.
Calibration Procedure Using Prepared Standards
Note: Deionized water or a reagent blank can be used to zero during the calibration
procedure. Calibrations generated with deionized water as the zero will give
less accurate results if the reagent blank is significantly more turbid or colored
than deionized water. Use the deionized water or the reagent blank as the zero
concentration point (S0) in the following calibration procedure.
1. Turn on the instrument and select the range to be calibrated. An arrow at
the top of the display will point to the selected range. To change ranges,
press the
MENU key, then use the READ/ENTER key to toggle between
ranges 1 and 2. Press
MENU again to return to measurement mode.
5957088.book Page 17 Friday, March 22, 2013 3:54 PM

Other manuals for Hach POCKET COLORIMETER II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach POCKET COLORIMETER II and is the answer not in the manual?

Hach POCKET COLORIMETER II Specifications

General IconGeneral
Measurement MethodColorimetric
DisplayLCD
Operating Temperature0 to 50 °C
Storage Temperature-40 to 60 °C
Waterproof RatingIP67
Weight0.23 kg
Light SourceLED
RangeVaries by parameter
WavelengthVaries by parameter
Battery LifeApproximately 2000 tests