2—20
User-Entered Calibration, continued
14. Press the READ/ENTER key. The meter will measure and display the
absorbance value for S1.
15. The calibration is complete with two points. If additional standards are
required, press
ZERO/SCROLL until “Add” appears on the display. Repeat
steps 1114 to enter additional standards.
16. Press the
MENU key twice to exit and accept the changes. The instrument
will use this calibration to determine the displayed concentration of future
sample measurements.
Entering a Predetermined Calibration Curve
Note: Entering a predetermined calibration curve requires at least two data pairs.
Each data pair requires a concentration value and the absorbance value for
the given concentration. Up to 10 data pairs may be entered. This procedure
uses the Edit mode.
1. Turn on the instrument and select the range to be calibrated. An arrow at
the top of the display will point to the selected range. To change ranges,
press the
MENU key, then use the READ/ENTER key to toggle between
ranges 1 and 2. Press
MENU again to return to measurement mode.
5957088.book Page 20 Friday, March 22, 2013 3:54 PM

Other manuals for Hach POCKET COLORIMETER II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach POCKET COLORIMETER II and is the answer not in the manual?

Hach POCKET COLORIMETER II Specifications

General IconGeneral
Measurement MethodColorimetric
DisplayLCD
Operating Temperature0 to 50 °C
Storage Temperature-40 to 60 °C
Waterproof RatingIP67
Weight0.23 kg
Light SourceLED
RangeVaries by parameter
WavelengthVaries by parameter
Battery LifeApproximately 2000 tests