EasyManua.ls Logo

Hach POLYMETRON 8810 - 7.1 System configuration (PM XXXX option)

Hach POLYMETRON 8810
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STOP - Pressing Enter causes the measurement process to stop immediately
START - Pressing Enter initiates the measurement process
STAND-BY - Pressing Enter puts the analyzer in standby
PROGRAMMING - Program the analyzer (Command mode only)
READING - View parameters (Operating mode only)
In PROGRAMMING or READING mode the following options can be selected using the scroll keys:
CALIBRATION - Calibrate the analyzer
ANALOG-OUT - Analog output parameters
ALARM - Alarm limit parameters
SEQUENCE - Measurement mode
SERVICE - Check the analyzer and accessory operations
CLEANING - Chemical cleaning parameters
SAMPLE COND. - Sample conditioning parameters
CAL PARAMETER - View the primary and last calibration details
TIMING - Titration cycle parameters
PM XXXX - System configuration (only available in PROGRAMMING mode and requires a valid
password)
Note: Some of the above options are only available if the relevant accessory has been installed.
7.1 System configuration (PM XXXX option)
Refer also to any Installation Procedure documents delivered with the analyzer for application
specific information.
1. Select PM XXXX with the scrolling keys and press Enter. Using the Select key to move from one
digit to the next, enter the special code and press Enter.
Option Description
ELECTRODE Define the electrode parameters:
Name
Slope
(Uiso) Isothermal point potential
(E) Reference point potential adjusted to 25°C
(C) Reference point concentration adjusted to 25°C
The E and C parameters are used to calculate the standard potential (E
025
= E -
S.logC)
UNIT Select the unit of concentration from the list available. Selecting USER allows you to
define your own, and requires a 5 character description.
Adjust ADC This option is reserved for Hach Lange service technicians.
CAL
TEMPERATURE
Two-point temperature calibration. Use a temperature simulator for each of the two
points. Wait until the measurement stabilizes before accepting each calibration point.
Adjust Iout Select the output type (0-20mA or 4-20mA). Place a multimeter in series with the
analog output and make any small adjustments to the signal as required.
20 English

Table of Contents

Other manuals for Hach POLYMETRON 8810

Related product manuals