5.4.1.1 Cable preparation
W A R N I N G
To conform with security standards and to respect the EMC compliancy of the analyzer, this procedure must be
followed for all user supplied cables before connecting them to the electronic unit of the analyzer.
Materials required:
Shielded cable (diameter minimum 4.5 mm - maximum 6 mm) with 2 or 3 wires depending on
function (RS232 = 3 wires, Alarm = 2 wires, etc.)
Metal cable gland
Grounding ferrule
2 or 3 protective plugs for the exposed wires
Clamp for the grounding ferrule
Crimping pliers for the protective plugs
English 9