3. Define the output:
Option Description
OUT: Conc Output corresponds to the last measured values.
OUT: Full Output is set to 20 mA.
OUT: Zero Output is set to 0 or 4 mA.
Note: You can only exit this menu when the option Conc is displayed.
7.4 Set the alarms and thresholds (ALARM option)
The analyzer has three alarms which are allocated according to the definition of WARNING/R1:
WARNING/R1 set to NO WARNING/R1 set to YES
System alarm System and warning alarm System alarm only
Alarm 1 Alarm 1 Warning alarm
Alarm 2 Alarm 2 Alarm 1 and 2
The alarms are allocated to the channel depending on the number of channels in operation:
2 Channels = NO 2 Channels = YES
Alarm 1
Channel 1
Channel 1
Alarm 2 Channel 2
1. Select ALARM with the scrolling keys and press Enter.
Option Description
SYS.ALARM ON/OFF Activate or deactivate the system alarm.
WARNING/R1 YES/NO Define the alarm allocation (refer to table above).
ALARM 1/2 ON/OFF Activate or deactivate alarms 1 and 2.
ALARM 1/2 HIGH/LOW If activated, define if the trigger is a high or low measurement value.
ALARM 1/2 XXX Enter the threshold value. If the alarm is set to HIGH any measured value above
this threshold will trigger the alarm, or if the alarm is set to LOW any measured
value below this threshold will trigger the alarm.
7.5 Measurement mode (SEQUENCE option)
Define the measurement mode.