Table of Contents
1 Online user manual on page 3
2 Specifications on page 3
3 General information on page 4
4 Installation on page 6
5 User interface and navigation on page 8
6 Startup on page 10
7 Standard operation on page 11
8 Calibration on page 15
9 Data log on page 16
10 Maintenance on page 17
11 Troubleshooting on page 20
12 Replacement parts and accessories on page 22
Section 1 Online user manual
This Basic User Manual contains less information than the User Manual, which is available on the
manufacturer's website.
Section 2 Specifications
Specifications are subject to change without notice.
Specification Details
Dimensions (W x D x H) 13.08 x 5.89 x 25.83 cm (5.15 x 2.32 x 10.17 in.)
Enclosure rating IP64 with caps installed for the meter probe connector ports or with
the probes connected to the instrument
Weight 1.2 kg (2.7 lb)
Power requirements (internal) Lithium ion rechargeable battery, 7.4V, 5.0Ah (5000mAh)
Power requirements (external) 100V–240 VAC, 50/60Hz input, 12V at 3A output
Operating temperature (discharging) 5 to 50 °C (41 to 122 °F), maximum 85% relative humidity (non-
condensing)
Charging temperature 5 to 45 °C (41 to 113 °F), maximum 85% relative humidity (non-
condensing)
Storage temperature –20 to 60 °C (–4 to 140 °F), maximum 85% relative humidity (non-
condensing)
Interface Mini USB port
5-pin input connector Two M12 connectors for IntelliCal probes
External power supply connection 2.5 mm Jack
Chemkey slots 4
Data memory 1000 measured values (result, date, time, site ID, user ID)
Data storage Automatic in Press to Read measurement mode. Manual in
Continuous measurement mode
Temperature correction (for IntelliCal
™
probes)
Off, automatic and manual (parameter dependent)
Measurement display lock (for IntelliCal
™
probes)
Press to Read and Continuous measurement mode
Protection class III
Certifications CE, UKCA, FCC, ISED, RCM, KC
Warranty 1 year
English 3