EasyManua.ls Logo

Hach TitraLab AT1000 Series - Set the Legacy Adapter Settings; Connect the Sensor

Hach TitraLab AT1000 Series
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 Temperature sensor 2 Reference sensor 3 Measuring sensor
3. The legacy adapter commissioning wizard automatically starts. Follow the instructions
that show on the display. Select the parameter based on the connected sensor.
Option Description
pH Select this parameter if the connected sensor is a pH analogic sensor.
Metal/RedOX/Color Select this parameter if the connected sensor is a Pt-Pt (metallic)
analogic sensor or PTM450/OPT300 sensor.
ISE Select this parameter if the connected sensor is an ion selective
sensor.
Refer to the application documentation to enter the related information for the
selected parameter.
3.8.1.1 Set the legacy adapter settings
Do the steps that follow to set a sensor name in the legacy adapter settings or set the
legacy adapter settings.
1. Make sure that the legacy adapter is connected to the instrument.
2. From the home screen, select Settings>Legacy settings and then push Select.
3. If more than one legacy adapter is connected, select the legacy adapter to change
the settings.
4. Select an option
2
. Refer to the application notes for additional information.
Option Description
Input sensor name Enters the name of the sensor used. If specified, follow the
recommendations on the application notes for the sensor name.
Input manual
temperature
Sets the temperature used in the titration if no temperature sensor
is connected.
Input pHIso (pH parameter only) Sets the pH ISO value (default=7).
Reset settings Sets the legacy adapter settings to the default configuration. Starts
the legacy adapter commissioning wizard.
3.8.2 Connect the sensor
Use a conical adapter to hold the sensor tightly in the sensor holder.
2
The options shown are based on the parameter selected on the legacy adapter commisioning wizard.
Installation
15