Section 4 Install accessories
4.1 Install an external pump
Table 3 External pump specifications (subject to change without notice)
Specification Details
Dimensions (W x D x H) 11.4 x 11.4 x 10.0 cm (4.49 x 4.49 x 3.94 in.)
Weight 280 g (0.62 lb)
Power requirements 12 VDC, 0.4–0.54 A
Altitude 2,000 m (6,562 ft) maximum
Operating temperature 15 to 35 °C (59 to 95 °F)
Relative humidity 20 to 80%, non-condensing
Storage temperature –5 to 40 °C (23 to 104 °F)
Installation category II
Pollution degree 2
Certifications EMC IEC/EN 61326-1
Warranty 1 year (EU: 2 years)
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this
product including, without limitation, direct, incidental and consequential damages, and disclaims
such damages to the full extent permitted under applicable law. The user is solely responsible to
identify critical application risks and install appropriate mechanisms to protect processes during a
possible equipment malfunction.
Please read this entire section before unpacking, setting up or operating this equipment.
Pay attention to all danger and caution statements. Failure to do so could result in serious
injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or
install this equipment in any manner other than that specified in this manual.
Read all labels and tags attached to the equipment. Personal injury or damage to the
equipment could occur if not observed. A symbol on the equipment is referenced in the
manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal
systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
Products marked with this symbol indicates that the product contains toxic or hazardous substances or elements. The
number inside the symbol indicates the environmental protection use period in years.
This accessory (item no. LZE142) is used to remove any excess sample from the beaker
to control the volume of sample before titration. The inlet and outlet tubes are pre-
installed on the pump. Make sure that the tube from the outlet port of the pump is
installed in an applicable liquid waste container. A blue tag is attached to the inlet tubes
that extract the liquid from the beaker.
To install the pump, refer to the illustrated steps that follow. After installation, check that
the pump and tubes are correctly installed by selecting Maintenance > Pump activation
from the instrument home screen (refer to Pump activation on page 43).