Questions and Answers

  • R
    Ronald CabreraAug 4, 2025
    What to do if Haefely Technology AXOS5 Test Equipment shows EUT requires more than 16 A?
    • L
      Laurie PattersonAug 4, 2025
      If your EUT requires more than 16A support current, it might be defective or have a short circuit. First, check the Inominal of the EUT, noting that the AXOS5 supports a maximum of 16A. Control the EUT independently to check for a short circuit. If a higher Inominal is required, an external CDN is necessary.
  • T
    Timothy GallegosAug 11, 2025
    Why does my Haefely Technology AXOS5 Test Equipment beep once and automatically shut down?
    • B
      Bridget MendozaAug 11, 2025
      If the AXOS5 Test Equipment produces one beep and automatically shuts down, it could be due to several reasons. It could be that at system startup any of the internal supply voltages (excluding 24V) are out of range, or the internal supply voltage 24V is out of range. Another cause might be a bad or missing Function Key or a missing or damaged internal storage card. In the first two cases send AXOS5 for repair. In the third case, go to Setup License Manager and check for correct numbers or send for repair.
  • M
    Manuel EverettAug 18, 2025
    What to do if I tried to start a Test with open Safety Circuit on Haefely Technology AXOS5?
    • X
      xshahAug 18, 2025
      If you tried to start a test with an open safety circuit on your Haefely Technology Test Equipment, close the safety circuit and start the test again.
  • M
    meganmorganAug 26, 2025
    What to do if Frequency of EUT Supply Input is too high on Haefely Technology AXOS5 Test Equipment?
    • T
      Tyler JohnsonAug 27, 2025
      If the frequency of the EUT Supply Input is too high on your Haefely Technology Test Equipment, use an EUT Supply Input voltage with a nominal frequency of 50 or 60 Hz.

Summary

1 SAFETY

1.1 General Safety Information

Crucial safety precautions and warnings for operating the AXOS.

2 PRODUCT DESCRIPTION

4 INITIAL OPERATION

4.1 Visual check

Instructions for inspecting the unit for damage before operation.

5 GENERAL OPERATION

6 SURGE

7 MAGNETIC FIELD

8 ELECTRICAL FAST TRANSIENT BURST

9 VOLTAGE DIPS & INTERRUPTS

10 ERROR MESSAGES

10.2 Troubleshooting

Common error messages and their possible causes and actions.

12 SERVICE

12.1 Verification

Procedures and equipment for verifying the AXOS performance.

Haefely Technology AXOS5 Specifications

General IconGeneral
BrandHaefely Technology
ModelAXOS5
CategoryTest Equipment
LanguageEnglish