6.3.5 Synchronization AC
To synchronize the surge signal with the main supply source it is necessary to adjust in the menu “Synchronization”
and then “Line Sync.”. Further adjustments of the angle can be done in a range from 0° to 359°. However, if no power
supply is connected to the EUT input at the rear view, it has to be entered “Async.”, otherwise it is impossible to get a
Surge impulse of the generator.
In external synchronization, the output with the position 8 Figure 5-2 in the rear view is in use. As an external source
could be used for instance a manual CDN. Normal operation through the touch screen is possible.