20 AXOS
5
Compact immunity test System
In Figure 7-3 is the Surge impulse proceeding. The magnetic field will be displayed and the safety symbol is flashing.
The test can be stopped by pushing the “Stop” button. When pushing the “home” button the starting screen as shown
in Figure 2-1 gets displayed.
Figure 7-3: Magnetic field charging
In following points “Coil Antenna”, “Trigger”, “Transition” and “properties” will be described. When being in the “Setup”
menu Figure 5-3, manual has been selected for “rep data”, therefore a test report can be written on a USB device
simply by pressing “Rep.Data”. However, first connect device to USB port.
7.1.1 Coil antenna
When the AXOS5 has started, the setup menu contains a standard predefined coil in accordance to IEC 61000-4-9. It
can be up to 10 different coil factors added to the menu.
Figure 7-4: Coil Antenna factor
To add a different coil, it will be entered in setup (Figure 5-3). This selection can be seen in the operating menu in the
magnetic field.