32 AXOS
5
Compact immunity test System
10.3
Logic errors
Alarm message Possible cause Action needed
Too many spikes per second
Decrease peak voltage / rep.
frequency / Burst or increase
burst period
Repetition frequency is too high
Decrease peak voltage ,
repetition frequency
burst period / burst duration incompatible
Decrease burst duration / Sync.
time
Incompatible settings
Decrease burst duration /Sync.
Time increase Burst period
Incompatible settings
Increase burst period, repetition
Frequency
Too many spikes per packet
Decrease Burst duration, Rep.
Frequency, Increase Burst
period
Incompatible settings for pre-compliance
mode
Burst mode must be set to
Normal, Trigger must be set to
Auto, Transition must be set off
Incompatible settings
Decrease peak voltage,
increase repetition rate
Incompatible settings
Decrease field strength,
increase repetition rate
Incompatible settings
Decrease dip duration or
increase dip interval
Entered value out of range
Value must be in the range of
min. and max. value Note:
Similar error can occur, only
with different units, however,
similar action for solving
Table 10-2: Logic causes and actions