EasyManua.ls Logo

Haefely Technology AXOS5 - SETUP Menu

Haefely Technology AXOS5
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 AXOS
5
Compact immunity test System
AUX (Pos. 7.) function description
1 Red warning lamp Indicates when safety circuit closed
2 Green warning lamp Indicates when safety circuit opened
3 GND 0 V
4 GND 0 V
5 Safety interlock When safety barriers installed, connection
must be done between pin 5 and pin 4 to close
the circle (safety instruments), however, bridge
between pin 4 and pin 5 exist as delivery
setting
6 GND 0 V
7 Start/Stop program 1 is running (Input) Start/Stop automatic for starting a predefined
program in the “Setup” menu “start file (Figure
5-3)”, signal for instance external device, PLC
signal
8 GND 0 V
9 Trigger Input Input for external trigger to start for instance
through an external signal a generator
10 GND 0 V
11 EUT Fail (Input) External condition (e.g. device or PLC signal),
the position becomes 1 and causes an action
(Ignore, Alarm, Test Stop or test stop & line off)
12 GND 0 V
13 0-10V Analogue output Reserved and used for further application
14 GND 0 V
15 Trigger output When Burst, Surge or Voltage dips package in
use, signal becomes 0, after the applied
package has finished the signal becomes
1(negation); used for indication of signal on
oscilloscope
16 GND 0 V
17 Spare (Reserve) Not in use
18 GND 0 V
Table 5-3: AUX function description
5.3
SETUP menu
The Figure 5-3 shows the “Setup” menu after selecting this button in the starting field. Every button is explained in the
Table 5-4.
Figure 5-3: Setup menu
In the right upper corner the current software version, serial number can be seen. The FPGA indicates the hardware of
the chip version.

Table of Contents