Mark this position on the thicker white band on
the top of the cuvet with a water resistant pencil.
Always use this position to align it with the
sign on the instrument top.
In order to match multiple cuvets follow the next
steps:
Fill some cuvets with high quality water
(<0.1NTU) up to the 10 mL mark.
Clean and oil the cuvets as described before.
Turn the instrument ON.
Insert the first cuvet into the instrument and
press “Read” functional key. Record the
reading.
Mark this position on the thicker white band on
the top of the cuvet with a water resistant pencil.
MATCHING MULTIPLE CUVETS
Precise measurements require the use of a single cuvet. If it is not possible, cuvet selection and
matching must be performed before taking measurements.