The HI 83414 offers an interactive context help mode that assists the user at any moment.
To access help screens, just press HELP.
Depending which menu you are in, a screen with
additional information will appear. To read all
available data, scroll down or up the text using
the UP or DOWN keys.
Press the “Support” functional key to access a page
with Hanna support centers and their contact details.
Press the “Accessories” functional key to access a
page with instrument accessories.
To exit support or accessories screens, press ESC, and
the instrument will return to the previous help screen.
To exit help mode just press the HELP key again and the meter will display the last screen the user
was in before entering help mode.
MEASUREMENT PROCEDURE FOR RATIO /MEASUREMENT PROCEDURE FOR RATIO /
MEASUREMENT PROCEDURE FOR RATIO /MEASUREMENT PROCEDURE FOR RATIO /
MEASUREMENT PROCEDURE FOR RATIO /
NON NON
NON NON
NON
RATIORATIO
RATIORATIO
RATIO
TURBIDITY RANGETURBIDITY RANGE
TURBIDITY RANGETURBIDITY RANGE
TURBIDITY RANGE
When taking turbidity measurements, several basic rules should be followed:
• Always use cuvets without scratches or cracks because they can cause inaccurate readings.
• Always cap the cuvets to avoid spillage of the sample into the instrument.
• Always close the lid of the instrument during measurement.
• Do not use too much oil to prevent contamination of the optical system.
To take turbidity measurements, follow the next steps:
• Turn the instrument ON by pressing ON/OFF.
When dashes are displayed on the LCD, the
instrument is ready. The current time appears on
the upper left corner of the display and the
range name appears on the lower right corner.
• Fill a clean, dry cuvet with 10 mL of sample up to the
mark, taking care to handle the cuvet by the top.
• Replace the cap.