EasyManua.ls Logo

Hanna Instruments HI83300 - Page 107

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
107
HYDRAZINE
# 2
# 1
• Place the cuvette (#1) into the holder and close the lid.
• Press Timer and the display will show the countdown prior to zeroing the blank or, alternatively
wait for 12 minutes and press Zero. The display will show “‑0.0‑” when the meter is zeroed
and ready for measurement.
• Remove the blank.
• Insert the cuvette with the reacted sample (#2) into the instrument
and close the lid.
• Press Read to start the reading. The instrument displays concentration in µg/L of hydrazine (N
2
H
4
).
INTERFERENCES
Interference may be caused by: highly colored samples, highly turbid samples, aromatic amines.

Table of Contents

Related product manuals