EasyManua.ls Logo

Hanna Instruments HI83300 - Zinc Measurement

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
188
8.60. ZINC
SPECIFICATIONS
Range 0.00 to 3.00 mg/L (as Zn)
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading at 25 °C
Light Source LED with narrow band interference filter @ 575 nm
Method Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18th edition, Zincon method.
REQUIRED REAGENT
Code Description Quantity
HI93731A‑0 Zinc Reagent A 1 packet
HI93731B‑0 Zinc Reagent B 0.5 mL
REAGENT SETS
HI93731‑01 Reagents for 100 tests
HI93731‑03 Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Zinc method using the procedure described in the Method
Selection section (see page 18).
• Fill the graduated glass vial up to the 20 mL mark with the sample.
• Add one packet of HI93731A‑0 Zinc Reagent A, close the
cylinder, and invert several times to mix until completely
dissolved.
• Fill a cuvette with 10 mL of the reacted sample (up to the
mark) and close the cap.
20 mL
10 mL