8.39. NICKEL HIGH RANGE
SPECIFICATIONS
Range 0.00 to 7.00 g/L (as Ni)
Resolution 0.01 g/L
Accuracy ±0.07g/L ±4% of reading at 25 °C
Light Source LED with narrow band interference filter @ 575 nm
Method Adaptation of the photometric method.
REQUIRED REAGENTS
Code Description Quantity
HI93726‑0 Nickel High Range Reagent 1 packet
REAGENT SETS
HI93726‑01 Reagents for 100 tests
HI93726‑03 Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
N
ickel HR method using the procedure described in the
Method Selection section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
10 mL
NICKEL HIGH RANGE