8.38. NICKEL LOW RANGE
SPECIFICATIONS
Range 0.000 to 1.000 mg/L (as Ni)
Resolution 0.001 mg/L
Accuracy ±0.010 mg/L ±7% of reading at 25 °C
Light Source LED with narrow band interference filter @ 575 nm
Method Adaptation of the PAN method.
REQUIRED REAGENTS
Code Description Quantity
HI93740A‑0 Nickel Low Range Reagent A 2 packets
HI93740B‑0 Nickel Low Range Reagent B 2 mL
HI93740C‑0 Nickel Low Range Reagent C 2 packets
HI93703‑51 Dispersing Agent (optional reagent) 4‑6 drops
REAGENT SETS
HI93740‑01 Reagents for 50 tests
HI93740‑03 Reagents for 150 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Nickel LR method using the procedure described in the
Method Selection section (see page 18).
Note: For best results samples should be between 20 and 24 °C.
• Fill one graduated beaker with 25 mL of deionized water (blank)
and another one with 25 mL of sample.
• Add one packet of HI93740A‑0 Nickel Low Range Reagent A to each
beaker. Cap and swirl gently until the reagent is dissolved.
Note: If sample contains iron (Fe
3+
), it is important that all
powder is dissolved before continuing.
• Add 1 mL of HI93740B‑0 Nickel Low Range Reagent B to each
beaker, and swirl to mix.