EasyManua.ls Logo

Hanna Instruments HI83300 - Page 125

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
125
MOLYBDENUM
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 5 minutes and press Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in mg/L of molybdenum (Mo
6+
).
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result to mg/L of molybdate (MoO
4
2-
) and sodium
molybdate (Na
2
MoO
4
).
• Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Aluminum above 50 mg/L
Chromium above 1000 mg/L
Copper above 10 mg/L
Iron above 50 mg/L
Nickel above 50 mg/L
Nitrite, as NO
2
¯
Sulfate above 200 mg/L
Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents

Table of Contents