EasyManua.ls Logo

Hanna Instruments HI83300 - Phosphate High Range

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
166
8.53. PHOSPHATE HIGH RANGE
SPECIFICATIONS
Range 0.0 to 30.0 mg/L (as PO
4
3‑
)
Resolution 0.1 mg/L
Accuracy ±1.0 mg/L ±4% of reading at 25 °C
Light Source LED with narrow band interference filter @ 525 nm
Method Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18th edition, Amino Acid method.
REQUIRED REAGENTS
Code Description Quantity
HI93717A‑0 Phosphate High Range Reagent A 10 drops
HI93717B‑0 Phosphate High Range Reagent B 1 packet
REAGENT SETS
HI93717‑01 Reagents for 100 tests
HI93717‑03 Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Phosphate HR method using the
procedure described in the Method Selection
section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for measurement.
• Add 10 drops of HI93717A‑0 Phosphate High Range Reagent A.
10 mL
PHOSPHATE HIGH RANGE