EasyManua.ls Logo

Hanna Instruments HI83300 - Alkalinity, Marine

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
10 mL
8.2. ALKALINITY, MARINE
SPECIFICATIONS
Range 0 to 300 mg/L (as CaCO
3
)
Resolution 1 mg/L
Accuracy ±5 mg/L ±5% of reading at 25 °C
Light Source LED with narrow band interference filter @ 610 nm
Method Colorimetric Method.
REQUIRED REAGENTS
Code Description Quantity
HI755S Alkalinity Reagent 1 mL
REAGENT SETS
HI755‑26 Reagents for 25 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Alkalinity Marine method using the procedure described in
the Method Selection section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
ALKALINITY, MARINE