84
8.22. CYANURIC ACID
SPECIFICATIONS
Range 0 to 80 mg/L (as CYA)
Resolution 1 mg/L
Accuracy ±1 mg/L ±15% of reading at 25 °C
Light Source LED with narrow band interference filter @ 525 nm
Method Adaptation of the turbidimetric method.
REQUIRED REAGENTS
Code Description Quantity
HI93722‑0 Cyanuric Acid Reagent 1 packet
REAGENT SETS
HI93722‑01 Reagents for 100 tests
HI93722‑03 Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Cyanuric Acid method using the procedure described in the
Method Selection section (see page 18).
• Fill the first cuvette with 10 mL of unreacted
sample (up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
# 1
10 mL
# 1
CYANURIC ACID