EasyManua.ls Logo

Hanna Instruments HI83300 - Ammonia High Range

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
AMMONIA HIGH RANGE
8.6. AMMONIA HIGH RANGE
SPECIFICATIONS
Range 0.0 to 100.0 mg/L (as NH
3
‑N)
Resolution 0.1 mg/L
Accuracy ±0.5 mg/L ±5% of reading at 25 °C
Light Source LED with narrow band interference filter @ 420 nm
Method Adaptation of the ASTM Manual of Water and Environmental Technology,
D1426, Nessler method.
REQUIRED REAGENTS
Code Description Quantity
HI93733A‑0 Ammonia High Range Reagent A 4 drops
HI93733B‑0 Ammonia High Range Reagent B 9 mL
REAGENT SETS
HI93733‑01 Reagents for 100 tests
HI93733‑03 Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Ammonia HR method using the procedure described in the
Method Selection section (see page 18).
• Add 1mL of unreacted sample to the cuvette using 1mL
syringe.
• Use the pipette to fill the cuvette up to the 10 mL mark
with HI93733B‑0 Ammonia High Range Reagent B.
Replace the cap and mix the solution.
• Place the cuvette into the holder and close the lid.
1 mL

Table of Contents

Related product manuals