48
8.7. BROMINE
SPECIFICATIONS
Range 0.00 to 8.00 mg/L (as Br
2
)
Resolution 0.01 mg/L
Accuracy ±0.08 mg/L ±3% of reading at 25 °C
Light Source LED with narrow band interference filter @ 525 nm
Method Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18th edition, DPD method.
REQUIRED REAGENTS
Code Description Quantity
HI93716‑0 Bromine Reagent 1 packet
REAGENT SETS
HI93716‑01 Reagents for 100 tests
HI93716‑03 Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Bromine method using the procedure described in the
Method Selection section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
10 mL
BROMINE