8.37. MOLYBDENUM
SPECIFICATIONS
Range 0.0 to 40.0 mg/L (as Mo
6+
)
Resolution 0.1 mg/L
Accuracy ±0.3 mg/L ±5% of reading at 25 °C
Light Source LED with narrow band interference filter @ 420 nm
Method Adaptation of the mercaptoacetic acid method.
REQUIRED REAGENTS
Code Description Quantity
HI93730A‑0 Molybdenum Reagent A 1 packet
HI93730B‑0 Molybdenum Reagent B 1 packet
HI93730C‑0 Molybdenum Reagent C 1 packet
REAGENT SETS
HI93730‑01 Reagents for 100 tests
HI93730‑03 Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Molybdenum method using the procedure described in the
Method Selection section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
10 mL
MOLYBDENUM