EasyManua.ls Logo

Hanna Instruments HI83300 - Calcium, Marine

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53
10 mL
8.9. CALCIUM, MARINE
SPECIFICATIONS
Range 200 to 600 mg/L (as Ca
2+
)
Resolution 1 mg/L
Accuracy ±6% of reading at 25 °C
Light Source LED with narrow band interference filter @ 610 nm
Method Adaptation of the Zincon method.
REQUIRED REAGENTS
Code Description Quantity
HI7581 Calcium Reagent A 1 mL
HI7582 Calcium Reagent B 1 packet
REAGENT SETS
HI758‑26 Reagents for 25 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Calcium Marine method using the procedure described in
the Method Selection section (see page 18).
• Add 1 mL of HI7581 Calcium Reagent A to the cuvette
using a 1 mL syringe.
• Use the plastic pipette to fill the cuvette to the 10 mL mark
with deionized water and replace the cap. Invert 3‑5 times
to mix.
• Place the cuvette into the holder and close the lid.
CALCIUM, MARINE

Table of Contents

Related product manuals