EasyManua.ls Logo

Hanna Instruments HI83300 - Iodine

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
108
8.31. IODINE
SPECIFICATIONS
Range 0.0 to 12.5 mg/L (as I
2
)
Resolution 0.1 mg/L
Accuracy ±0.1 mg/L ±5% of reading at 25 °C
Light Source LED with narrow band interference filter @ 525 nm
Method Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18th edition, DPD method.
REQUIRED REAGENTS
Code Description Quantity
HI93718‑0 Iodine Reagent 1 packet
REAGENT SETS
HI93718‑01 Reagents for 100 tests
HI93718‑03 Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the Iodine method using the procedure described in the
Method Selection section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
10 mL
IODINE

Table of Contents

Related product manuals