Remove the cuvette and add one packet of HI93726‑0
Nickel High Range Reagent. Replace the cap and shake
gently until completely dissolved.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 1 minute and press Read. When the timer ends the meter will perform the reading. The
instrument displays the concentration in g/L of nickel (Ni).
INTERFERENCES
Interference may be caused by copper.
NICKEL HIGH RANGE