• Add 0.2 mL of the HI93748B‑0 Manganese Low Range
Reagent B to each cuvette, replace the caps and invert
gently to mix for about 30 seconds.
• Add 1 mL of the HI93748C‑0 Manganese Low Range
Reagent C to each cuvette, replace the caps and shake
gently.
• Add 3 drops of HI93703‑51 Dispersing Agent to
each cuvette, replace the caps and invert gently to
mix for about 30 seconds.
• Place the first cuvette (#1) with the reacted deionized
water into the holder and close the lid.
• Press Timer and the display will show the countdown prior to zeroing the blank. Alternatively
wait for 2 minutes and then press Zero. The display will show ”‑0.0‑” when the meter is zeroed
and ready for measurement.
MANGANESE LOW RANGE