148
OXYGEN SCAVENGERS (DEHA)
• Add 0.5 mL of HI96773B‑0 Oxygen Scavengers
Reagent B to each cuvette using the 1 mL syringe.
• Replace the caps and invert for 10 seconds.
• Place first cuvette (#1) into the holder and close the lid.
• Press Timer and the display will show countdown prior to the measurement or, alternatively,
wait for 10 minutes and press Zero. The display will show “‑0.0‑“ when the meter is zeroed
and ready for measurement.
#1
#2
#1
• Add one packet of HI96773A‑0 Oxygen Scavengers
Reagent A to (#2) cuvette. Replace the cap and
invert for 30 seconds.
#2