EasyManua.ls Logo

Hanna Instruments HI83300 - Page 163

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
163
PHOSPHATE, MARINE ULTRA LOW RANGE
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 3 minutes and press Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in µg/L of phosphorus (P).
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result to µg/L of phosphate (PO
4
3-
) and phosphorus
pentoxide (P
2
O
5
).
• Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Iron above 50 mg/L
Silica above 50 mg/L
Silicate above 10 mg/L
Copper above 10 mg/L
Hydrogen sulfide, arsenate, turbid sample and highly buffered samples also interfere.

Table of Contents

Related product manuals