EasyManua.ls Logo

Hanna Instruments HI83300 - Page 79

Hanna Instruments HI83300
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
79
COLOR OF WATER
• Fill the second cuvette (#2) up to the mark with unfiltered
sample and replace the cap. This is the apparent color.
• Filter 10 mL of sample through a filter with a 0.45 µm
membrane into the third cuvette (#3), up to the 10 mL
mark and replace the cap. This is the true color.
• Insert the apparent color cuvette (#2) into the instrument
and close the lid.
• Press Read to start the reading. The meter displays the value of apparent color in Platinum
Cobalt Units (PCU).
• Remove the apparent color cuvette(#2) from the
instrument, insert the true color cuvette (#3) into the
instrument and close the lid.
•Press Read to start the reading. The meter displays the true color in Platinum Cobalt Units
(PCU).
10 mL
#2
#3
#2
#3

Table of Contents