5-9
TITRATION METHODS
5.5.5. DOSING TYPE
Option: Linear or Dynamic
5.5.5.1. LINEAR DOSING
Linear dosing dispenses a pre-defined volume of titrant with every addition.
The Linear Dosing option is recommended for titrations with a slower reaction rate, difficult nonaqueous titrations, and specific
applications.
Note: For steep and normal titration curves, smaller volume increments are recommended, to obtain many points around the
equivalence point.
For flat titration curves, larger volume increments are recommended for equivalence point detection.
To set the dosing volume, select Linear Dosing and enter the optimum dose.
Dosing volume ranges are:
5 mL burette 0.001 to 4.750 mL
10 mL burette 0.001 to 9.500 mL
25 mL burette 0.005 to 23.750 mL
50 mL burette 0.005 to 47.500 mL