delta E: maximum change in potential during delta t
The range is from 0.1 to 99.9 mV.
delta t: the time interval during which the potential is measured.
The range is from 1 to 10 seconds.
t min wait: the minimum elapsed time before a stability check. This is also the minimum elapsed time between two doses.
The range is from 2 seconds to t max wait time.
t max wait: the maximum elapsed time between two successive doses. If the t max wait has elapsed, a new dose is added
even if the signal stability condition is not reached.
The range is from t min wait time to 180 seconds.
5.5.10.2. TIMED INCREMENT
Option: 2 to 180 seconds
When timed increment is selected, the titrator acquires the potential (mV) at a fixed time interval (no signal stability check).
The time period between two acquisitions must be set according to the reaction and the response time of the electrode.