41
Notes: These procedures calibrate the slope value. To
calibrate the offset, set the calibration point at 0 μS/cm
and repeat the procedure.
If the temperature input is not within the acceptable
range (0 to 50 °C), the message “Invalid temperature”
is displayed.
If the conductivity input is not within the acceptable range, the message
“Wrong standard” is displayed.
The probe is factory calibrated for temperature readings. The user can perform a single point tem
perature calibration or restore factory calibration.
This procedure requires a reference temperature measuring instrument.
• Select “Temperature” from the “Calibration” menu.
• Select “Calibrate temperature”.
• Insert the probe in an isothermal bath with reference instrument and allow the probe to come to
thermal equilibrium.
• Use the keypad to enter the known temperature and then press Accept to confirm.
• The stability timer will count down and the message “Ready
and “Confirm” will be displayed.
• Press Confirm to store the calibration point.
After confirmation, the following messages are displayed
“Storing” and “Calibration completed”.
• Press OK to return to the “Calibration” menu.
• Press Measure to return to the measurement screen.
To restore the factory calibration, select the corresponding
option in the “Temperature calib.” menu and then press
Select.
TEMPERATURE CALIBRATION

Other manuals for Hanna Instruments HI98196

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI98196 and is the answer not in the manual?

Hanna Instruments HI98196 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI98196
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals