Specifications
30 A03729_03_YI0_01 HBM: public C9C
Nominal (rated) force 20010050NF
nom
Nominal (rated) force
5020105210.5kN
F
nom
Temperature
Reference temperature t
.ref
°C 23
Nominal (rated)
temperature range
B
t,nom
°C 10…+70
Operating temperature
range
B
t,g
°C 30…+85
Storage temperature
range
B
t,S
°C 30…+85
Characteristic mechanical quantities
Maximum operating force
F
G
% of
F
nom
200 120
Force limit
F
L
% of
F
nom
>200 >150
Breaking force
F
B
% of
F
nom
>400
Permissible eccentricity
when loaded with nominal
(rated) force
e
g
mm 2.6 2.5 2.5 3.5 2.6 3.2 1.8 2.0 0.8 2.5
Nominal (rated)
displacement ±15%
S
nom
mm
0.009 0.015 0.019 0.020 0.025 0.040 0.055 0.075
Fundamental frequency f
G
kHz 7.3 10 15.7 3.5 5 7 13 15.1 20 12
Permissible oscillation
stress
F
rb
% of
F
nom
80 70
Maximum impact load to ICE 60068-2-6
Number 1,000
Duration ms 3
Acceleration m/s
2
1,000
Vibrational stress to IEC 60068-2-27
Frequency range Hz 5 … 65
Duration min 30
Acceleration m/s
2
150
General information
Degree of protection per
EN 60529
IP67
Spring element material Steel