Specifications
C9C A03729_03_YI0_01 HBM: public 29
9 Specifications
C9C force transducer
Nominal (rated) force F
nom
N 50 100 200
kN 0.5 1 2 5 10 20 50
Accuracy
Accuracy class 0.2
Relative reproducibility
and repeatability errors
with unchanging
mounting position
b
rg
% <0.2
Relative reversibility error v % <0.2
non-linearity d
lin
% <0.2
Relative creep
d
crf+E
% <0.2 <0.1
Temperature coefficient of sensitivity
In the nominal (rated)
temperature range
TC
S
%/10
K
<0.2
In the operating temperature
range
<0.50
Temperature coefficient of zero signal
In the nominal (rated)
temperature range
TC
0
%/10
K
<0.2
In the operating temperature
range
<0.50
Rated electrical output
Rated output (nominal) C
nom
mV/V 1
Relative zero signal error d
s,0
mV/V ± 0.2
Sensitivity error d
c
% <1
Input resistance R
e
Ω 250  400 300  450
Output resistance R
a
Ω 200  400 100  450
Insulation resistance R
is
Ω > 1*10
9
Operating range of the
excitation voltage
B
u,gt
V 0.5….12
Reference excitation
voltage
U
ref
V 5
Connection 4-wire circuit